Specials
451
•Interventional Window 1/2/3
With the special window toggle button you can apply
three different window settings to your image, the
default values are: Abdomen (300/40), Lung (1200/-
600) and Bone (1500/450).
You can choose different window settings under
Options>Configuration>Examination>Interven-
tion, these settings correspond to the window values
list.
•Save and Auto Stop functions
The icon for Auto-Stop at saved TP will be highlighted
in green, after you have pressed the button for save
current Table Position, you will find a new entry in the
Drop down menu of the subtask card. By moving the
table to another position the icon will be displayed in
gray until you reach the saved Table position, then it
will turn to green again. If you activate the Auto-Stop
at saved TP function and use the Joystick or the but-
tons on the Gantry, the table will stop automatically at
the desired position so it is easier for you to reposition
the patient again. This function will stay active until
you press the button Auto-Stop at saved TP again.
The same behavior happens if you use the Auto-Stop
at Last Interventional Scan Position function. If you
press Auto-Stop at Last Interventional Scan Position
and use the Gantry buttons to position your patient,
the Gantry buttons will flash and show you in which
direction you need to move the table to get back to
your interventional table position again.