424
Specials
TraumaThis is a one-range mode for fast screening for emer-
gency studies.
For SOMATOM Emotion 16-slice configuration:
A scan range of 50 cm will be done in 11.62 s.
Emotion 16 Trauma 2nd
recon.
3rd
recon.
kV 130
Effective mAs/
Quality ref. mAs
85
Rotation Time 0.6 sec.
Acquisition 16 x 1.2 mm
Slice collimation 1.2mm
Slice width 8.0mm 2. 0 mm 2.0 mm
Feed/Rotation 28.8 mm
Pitch Factor 1.50
Increment 8.0 mm 1.5 mm 1.5 mm
Kernel B31s B31s B70s
CTDIVol 9.52mGy
Effective dose Male: 7.02 mSv
Female: 8.81 mSv