Head
197
Hint
• For low dose studies use only 20 mAs.
Emotion 6 Sinus 2nd
recon.
3d
recon.
kV 130
Effective mAs/
Quality ref. mAs
35
Rotation Time 1.0 sec.
Acquisition 6 x 1.0 mm
Slice collimation 1.0mm
Slice width 5.0mm 3.0mm 3.0 mm
Feed/Rotation 5.1 mm
Pitch Factor 0.85
Increment 5.0 mm 3.0 mm 3.0 mm
Kernel H70s H30s H60s
CTDIVol 9.63 mGy
Effective dose Male: 0.20 mSv
Female: 0.25 mSv
Contrast medium IV injection
Start delay 45 sec.
Flow rate 2.0ml/sec.
Total am ount 75ml