402
Vascular
BodyAngioVol
Indications:
Spiral mode for abdominal CT Angio studies.
Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in 3D
images display view. The coronal view images will be
reconstructed as MIP images.
For SOMATOM Emotion 16-slice configuration:
A typical study in a range of 40 cm will be covered in
17.87sec.
For SOMATOM Emotion 6-slice configuration:
A typical study in a range of 20 cm will be covered in
19.38sec.
Emotion 16 BodyAngio 2nd reconstr.
kV 110
Effective mAs/
Quality ref. mAs
90
Rotation Time 0.6 sec.
Acquisition 16 x 0.6 mm
Slice collimation 0.6 mm
Slice width 5.0 mm 3.0 mm
Feed/Rotation 14.4 mm
Pitch Factor 1.50
Increment 5.0mm 3.0 mm
Kernel B31s B30s
CTDIVol 7.20 mGy
Effective dose Male: 5.04 mSv
Female: 6.36mSv