Vascular
371
Hint
• Use of CARE Bolus with monitoring scans positioned
at the level of the basilar artery or carotid artery. Set
the trigger threshold at 120 HU, or use manual trig-
gering.
Emotion 6 HeadAngio 2nd reconstr.
kV 110
Effective mAs/
Quality ref. mAs
70
Rotation time 0.8 sec.
Acquisition 6 x 1.0mm
Slice collimation 1.0mm
Slice width 3.0 mm 4.0 mm
Feed/Rotation 5.1mm
Pitch Factor 0.85
Increment 3.0mm 4.0 mm
Kernel H31s H20s
CTDIVol 12.88m Gy
Effective dose Male: 0.37 mSv
Female: 0.42mSv
Contrast medium IV injection
Start delay 18 sec.
Flow rate 3.5ml/sec.
Total am ount 75ml