Thorax
245
Hints
• For lung cancer evaluation, this protocol can be com-
bined with protocol “Neck Routine”.
Emotion 6 ThorFast 2nd recon.
kV 130
Effective mAs/
Quality ref. mAs
70
Rotation Time 0.8/0.6sec.
Acquisition 6 x 3.0mm
Slice collimation 3.0mm
Slice width 6.0mm 6.0 mm
Feed/Rotation 27.0mm
Pitch Factor 1.50
Increment 6.0mm 6.0mm
Kernel B41s B70s
CTDIVol 7.14mGy
Effective dose Male: 4.12 mSv
Female: 5.25mSv
Contrast medium IV injection
Start delay 25 – 30 sec.
Flow rate 2.5ml/sec.
Total amount 80 ml