Children
589
Hints
• CARE Bolus may be used to optimize bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta with triggering threshold of 120 HU, or use
manual triggering.
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access
site
Total
amount
1 – 2 ml per kg of body weight