DISPlay Subsystem
3–64 AM700 Audio Measurement Set Programmer Manual
DISPlay[:WINDow[1–4]]:TRACe[1|2]:X[:SCALe]:RIGHt
<numeric_value>
Usage: Sets or queries the value represented by the maximum (right) edge of the x-axis.
This value may be bounded by a range of data. When a new RIGHT value is
entered, the divisions of the scale remain the same, but the CENTer and LEFT
values are changed.
Query: DISP:WIND1:TRAC1:X:SCAL:RIGH? returns the value of the right edge
of the x-axis scale of trace 1 in window 1.
*RST: At *RST, the RIGHT value is set to the maximum value available for the
default application, for example 20000 Hz in the FFT analyzer, high resolution
mode.
DISPlay[:WINDow[1–4]]:TRACe[1|2]:X:SPACing
<LINear|LOGarithmic>
Usage: Sets the horizontal scale to either linear or logarithmic display of the trace data.
Query: DISP:WIND1:TRAC1:X:SPAC? returns the type of x-axis scaling in use,
either LINEAR or LOGARITHMIC, for trace 1 of window 1.
*RST: Sets the Analyzer Wind[1–2] x-axis spacing to LINear, Wind[3–4] to LOG.
Monitor x-axis spacing to LINear.
DISPlay:WINDow:TRAce:Y:AXIS:UNIT <unit_type>
Usage: Used to set the units of the traces y-axes.
Example: DISP:WIND:TRAC:Y:AXIS:UNIT volts sets the trace y-axis to volts.
Query: DISP:WIND:TRAC:Y:AXIS:UNIT? returns the setting of the y-axis units.
*RST: Sets the y-axis units to dBu.