STATus Subsystem
3–144 AM700 Audio Measurement Set Programmer Manual
The following commands can be applied to all SCPI status registers by prefixing the command with
the node(s) that represent the particular register to be controlled.
STATus:OPERation

Table 3–4: Summary of STATus:OPERation

Summary bit settings Condition
0 Calibration
3 Sweep
5 Trigger
12 System
13 Instrument Summary
14 Program Running

Table 3–5: STATus:OPERation

Condition definition Condition number Status
Calibration 100 Set if calibration is occurring
Sweep 103 Set when a sweep is started;
cleared with a sweep is
finished
Trigger for Monitor and FFT 105 Status Bits:
3 Waiting for Mtone Ch 1
4 Waiting for Mtone Ch 2
5 Triggered Ch 1
6 Triggered Ch 2