CALibration Subsystem
3–40 AM700 Audio Measurement Set Programmer Manual
CALibration:OUTPut:POWer:REFerence:RESistance
<numeric_value>
Usage: Set the user’s external resistor values to use in the AM700 display of output
power (dBm).
Range: 0.001 to 200000.0
Resolution: 0.001
Units: Ohms (impedance)
*RST: No change
NOTE. Output dBm calculation takes both this reference resistance and
the generator source impedance into account.
CALibration:VOLTage:FS <numeric_value>
Usage: Sets or queries the rms voltage corresponding to 0 dBFS. This calibration
command is used so the digital generator can follow the analog generator, and
to enable analog and digital measurements to be overlayed in the same view.
Units: Volts
*RST: No change