CMODe Subsystem (FFT/MTONe Analyzer)
AM700 Audio Measurement Set Programmer Manual 3–45
CMODe Subsystem (FFT / MTONe Analyzer)
Usage: Commands under CMODe:FFT are used to switch the FFT Analyzer operations
between Zoom mode and Multitone measurements and to designate signal
sources for the multitone signal. The CMODe:FFT:WINDow commands permit
specifying a user specified window file.

CMODe:FFT FFT|MTONe

Usage: Selects either FFT (Zoom) or Multitone measurements for the FFT Analyzer.
Default: The power on default for CMODe is FFT.
Query: CMOD:FFT? returns the selection for the FFT Analyzer measurement as either
FFT or MTONe.
*RST: Sets CMODe to FFT.

CMODe:FFT:MTONe[1|2]:AVECtor?

Usage: Returns the amplitude of the individual tones used to generate the designated
multitone signal.

CMODe:FFT:MTONe[1|2]:FILE:NAME ’file_name’

Usage: Names the file used to generate the designated multitone when file mode is
selected.
Query: CMOD:FFT:MTON[1|2]:FILE:NAME? returns the file name of the file
selected to supply the data used to generate the designated multitone signal.
*RST: Filename is set to ”asgmton1.ton” for MTON1 and MTON2.

CMODe:FFT:MTONe[1|2]:FVECtor?

Usage: Returns the frequency of the individual tones used to generate the designated
multitone signal.