OUTPut Subsystem
3–92 AM700 Audio Measurement Set Programmer Manual
OUTPut:COMMon FLOat|GROund
Usage: Selects whether the analog outputs are floating (with no ground reference) or
referenced to ground.
Query: OUTP:COMM? returns “FLOAT” or “GROUND”.
*RST: Set the output common to FLOAT.
OUTPut3:FILTer[:LPASs][:STATe] <Boolean>
Usage: Turns the long cable simulation filter on in the output path.
Query: OUTPut:FILTer[:LPAss][:STATe]? returns 1 for ON and 0 for OFF.
*RST: Sets the long cable simulation filter state to OFF
OUTPut3:FILTer[:LPASs]:TYPE?
Usage: Query only that responds with CABLE. The query is in place to conform to
SCPI standard practices.