OUTPut Subsystem
3–90 AM700 Audio Measurement Set Programmer Manual
Parameters: frequency from 30 kHz to 52 kHz, with a resolution of 0.001 Hz.
Range: 30,000 to 52,000
Resolution: 0.001 Hz
Units: Hz
Query: OUTP3:DAUD:INT:CLOC:VFR? returns the variable clock frequency
setting.
*RST: 48000.0
OUTPut3:DAUDio:INTerface:CLOCk:FRequency:ADJust
<numeric_value>
Usage: Add a small offset to the digital interface clock frequency when
:CLOCk:MODe is set to AUDio or DARS.
Parameters: Frequency offset in parts per million
Query: OUTP3:DAUD:INT:CLOC:FREQ:ADJ? returns the offset setting.
Range: –100 to 100 ppm
Resolution: 1 ppm
*RST: 0 ppm
OUTPut3:DAUDio:INTerface:CLOCk:MODE AUDio|DARS|VARiable
Usage: Sets the clock mode to audio, digital audio reference signal, or variable. The
setting of mode determines which of the CLOCk frequency setting commands is
active.
Query: OUTP3:DAUD:INT:CLOC:MODE? return the mode setting.
*RST: AUDio