AMEasure Subsystem (Audio Analyzer)
3–10 AM700 Audio Measurement Set Programmer Manual
AMEasure[1–4]:XY:X:INPut <char_data>
Parameters: CHANnel1|CHANnel2|ANALog1|ANALog2|DIGital1|DIGital2
Usage: Sets or queries the source of the input to the x-axis for the designated AME in
XY mode.
Query: AME[1–4]:XY:X:INP? returns the measurement channel currently
supplying input to the designated AME.
*RST: AME1 CHANNEL1
AME2 CHANNEL2
AME3 CHANNEL1
AME4 CHANNEL1
AMEasure[1–4]:REGulation:COUNt <numeric_value>
Usage: Sets or queries the Sweep Count setting of the designated AME for the
regulation mode sweep count.
Range: 0 to 10000 counts.
Query: AME[1–4]:REG:COUN? returns the current setting for the regulation mode
sweep.
*RST: Sets count to 1.
AMEasure[1–4]:REGulation:ERRor <numeric_value>
Usage: Sets or queries the Target Error value of the designated AME for regulation
mode operation.
Range: Range of the numeric value depends on the setting of AME:REG:FUNC
<char_data>.
FREQ range = 0 to 81000
LEV range = –97.7815 to 47.0104
THD range = 0 to 100