Getting Started
1–22 AM700 Audio Measurement Set Programmer Manual
Syntax: ROUTe[1|2|3]:CLOSE <channel_list>
or
ROUTe[1|2|3]:CLOSE:STATE?
Only certain closures go with a given route: The acceptable routes and closures are:
rout1:clos 1|3
rout2:clos 2|4
rout3:clos 6|7|8|9
Example: The Close command allows specific individual channels to be closed or
queried. To select the generator as the signal source for the Analog A input, the
command is:
ROUT1:CLOS 3
The rout2[1|2|3]:close:state? query returns the <channel_list> number
for the designated route. If all the specified channels cannot be closed (trying to
close two channels at the same time if that is not allowed) an execution error is
reported. See the ROUTe Subsystem Command description in Section 3 for
complete information on the route command as used in the AM700.
The Input subsystem commands set the conditions prior to sampling. Here the
choices of input impedance and input range are made. The input choices are also
user selectable from within some measurement applications. The choices of
selecting the filtered output are not is a measurement choice that will be made as
needed by the application. Selections for the notch filter range and use are not
selectable via remote control.
Suffixes: Suffixes are: INP1 Analog acquisition channel ‘A’
INP3 Analog acquisition channel ‘B’
The syntax for the Input commands is as follows:
Syntax: INPut:RANGe 0.0870 to 173.616
INPUT:IMPedance 150|600|200000
INPut
Subsystem