CALCulate Subsystem (Average)
3–34 AM700 Audio Measurement Set Programmer Manual
Query: CALC[3|4]:AVER:TYPE? returns the current setting for the designated
CALC block as MAXIMUM, MINIMUM, or EXPONENTIAL.
*RST: Sets AVER:TYPE to EXPonential.
CALCulate[3|4]:FEED?
Usage: Query to determine the signal feed for the AVERage CALC blocks.
Query: CALC[3|4]:FEED? returns the feed for the designated CALC block. FFT
CALC3 is fed by ’CALC:TRAN:FREQ:MAG’, CALC4 is fed by
’CALC2:TRAN:FREQ:MAG’, and Jitter Spectrum CALC3 is fed by
’SENse 11’. These feeds cannot be changed.