ROUTe Subsystem

3–102 AM700 Audio Measurement Set Programmer Manual
ROUTe Subsystem
Signal routing is the block where the user has access to actual signals.

ROUTe[1–3]:CLOSe <numeric_value>

Usage: Sets the route setting. If all the specified channels cannot be closed, an
execution error is reported.
Suffixes: The ROUTe suffixes are as follows:
1 Analog A input
2 Analog B input
3 Digital main input
Parameters: Defined values for the CLOSe command <numeric_value>
1 analog A input connector
2 analog B input connector
3 analog generator channel A
4 analog generator channel B
6 digital front panel XLR
7 digital rear panel BNC
8 digital rear panel optical
9 AES internal generator
Only specific numeric_values are permitted for each route suffix. You can’t close a
route from the Analog A input to the Digital front panel XLR for instance. The
acceptable routes and closures are as follows:
rout1:clos 1|3
rout2:clos 2|4
rout3:clos 6|7|8|9
Example: Choose the generator for analog B input as follows:
ROUTe2:CLOSe 4
Query: ROUTe[1–3]:CLOSe? <channel_number>