Analog Programming (APG) and Isolated Analog Programming (ISOL)
4-10 M370046-01
Voltage-Controlled Voltage APG Setup

Activating APG Voltage Mode

To activate APG voltage mode using an external voltage source:
1. Turn the 9-position mode control to the VAP position or press the
rotary Adjust/Enter control if the mode control is already at the VAP
position.
UAPr is displayed on the output voltage display.
2. Turn the rotary Adjust/Enter control to select Un1S (voltage source
from non-isolated connector).
3. Press the rotary Adjust/Enter control to commit the selected mode.
LE U is displayed on the output voltage display.
4. Once the desired level has been selected, press the rotary Adjust/Enter
control to commit the change.
The display will return to the normal display.

Deactivating APG Voltage Mode

To deactivate the APG voltage mode:
1. Turn the 9-position mode control to the VAP position or press the
rotary Adjust/Enter control if the mode control is already at the VAP
position.
2. UAPr is displayed on the output voltage display.
3. Turn the rotary Adjust/Enter control to select nonE.
4. Press the rotary Adjust/Enter control to commit the change.
The display will return to the normal display.