Calibration and Troubleshooting
B-4 M370046-01
Step 1: Gain Calibration
Adjust the gain so that the real line and ideal line intersect at a
programmed value of 90%.
Step 2: Offset Calibration
Adjust the offset so that the real and ideal lines intersect at a programmed
value of 10%.
Figure B-2
Calibration: Step 1 Gain Calibration
Figure B-3
Calibration: Step 2 Offset Calibration