Local Operation

Table 3-9Power Supply Default Settings

Parameter

Setting

 

 

UVP

0 V

Foldback trigger

None

Foldback delay

0.5 s

Current Share Mode

MASTer

Alarm Output Latches

263 (0 × 107, all latches enabled)

Alarms Mask

2047 (0 × 7FF, all enabled)

Interlock

Disabled

Voltage Analog Programming

Off

Voltage APG Scale

10 V

Current Analog Programming

Off

Current APG Scale

10 V

 

 

Hard Reset

The hard reset performs the same functions as a soft reset, but will also

 

clear all calibration constants. Do not perform a hard reset unless you are

 

planning to re-calibrate all settings of your power supply.

 

To reset the power supply:

 

1.

Turn the front panel power switch to the Off position.

 

2.

Allow the unit to power down.

 

3.

When the fans are off and the front panel displays are not illuminated,

 

 

turn the power switch to the On position.

 

4.

Press and hold the OUTPUT ENABLE Main and OUTPUT

 

 

ENABLE Aux buttons.

 

 

The front panel should display P5U S rS.

 

5.

When the display reads P5u Clr, release the OUTPUT ENABLE

 

 

Main and OUTPUT ENABLE Aux buttons. After 1 second the model

 

 

number will appear on the display for 1 second.

The power supply has been reset.

3-40

M370046-01

Page 84
Image 84
Xantrex Technology XTR 850 WATT SERIES manual To reset the power supply