Calibration and Troubleshooting

Offset Calibration

Offset calibration of the power supply provides the best accuracy in low- range input current.

To perform offset calibration:

1.Set the voltage source attached to the non-isolated current APG input to 10% of the nominal current level, in this case 0.400 V.

2.Turn the output power on.

3.Read the input voltage value from the voltmeter connected in step 2 of the gain calibration.

4.Type SCPI input current calibration command with voltage noted in step 3 as the parameter “data”.

For example, you would type :CAL:INP:ANAL:CURR 0.43. After performing offset calibration, it is highly recommended that you repeat gain calibration.

SCPI Command [:]CALibration[<channel>]:INPut:ANALog:CURRent{<voltme ter reading in volts>}