Calibration and Troubleshooting
Non-isolated Resistive Programming of Voltage Calibration
Important: For maximum accuracy at a specific APG level. Follow the calibration procedure again but use the intended operating resistance level in steps 2 and 3.
To calibrate the
1.Disconnect the load and connect the voltmeter to the output.
2.Connect the 4.000kΩ resistor to the
3.Set the resistive APG level to 4kΩ. The SCPI command is:
SYST:COMM:APR:LEV 4
4.Set the input source as Resistive
SYST:REM:SOUR ARES
5.Turn the main output On by pressing the OUTPUT ENABLE Main button on the front panel or send the command:
:OUTP ON
The OUTPUT ENABLE Main button will illuminate.
6. Enter the interactive calibration mode by sending the SCPI command:
CAL:OUTP:ANAL:RES
7.Press the terminal keys – or + to reduce or increase main output voltage value. Adjust the main output voltage to 100%.
8.Press any key other than the + or – keys to exit from interactive calibration mode.