Status Byte

Standard Event Status Register (SESR)

The standard event status register sets bits for specific events during power supply operation. All bits in the standard event status registers are set through the error event queue. The register is defined by IEEE 488.2 register and is controlled using 488.2 common commands: *ESE, *ESE?, and *ESR? as well as SCPI aliases for multichannel use.

Standard Event Status Enable Register

The Event Summary Enable command determines which bits in the Standard Event Status Register are summarized in the Event Summary Bit (ESB) of the Status Byte. This register will be cleared at power up. To enable events, you must make the logical sum of all values for the events that you to be reported and send this parameter in decimal format.

See Table 5-9 for values for the events to be enabled in <status-enable> parameter.

For example, sending *ESE 16 sets bit 4 of the Standard Event Status Enable Register. This will cause the Event Summary bit (ESB) in the Status Byte to be set whenever the Execution Error bit (bit 4) in the Standard Event Status Register gets set.

Command:

*ESE <status-enable>, *ESE?

The SCPI equivalent for multichannel use:

5

[:]STATus[<channel(s)>]:STANdard:ENABle <status-enable> [:]STATus[<channel(s)>]:STANdard:ENABle?

Standard Event Status Register

The Standard Event Status Register query allows you to determine the current contents of the Standard Event Status Register. (See “Standard Event Status Register (SESR)”. Reading this register clears it.

Command:

*ESR?

The SCPI equivalent for multichannel use:

[:]STATus[<channel(s)>]:STANdard[:EVENt]?

M370046-01

5-31

Page 157
Image 157
Xantrex Technology XTR 850 WATT SERIES manual Standard Event Status Register Sesr, Standard Event Status Enable Register