Local Operation
3-14 M370046-01
Quick Tip
Remote operation
uses these SCPI
commands. See
page 5–1.
The SCPI command (s) for these instructions are:
[[:]SOURce[<channel>]]:VOLTage[:LEVEl][:IMMediate]
[:AMPLitude]
To set the current set point (ISET):
1. Turn the 9-position mode control to the AMPS position or press the
rotary Adjust/Enter control if the 9-position mode control is already
at the AMPS position.
The voltage set point is blinking in the output voltage display. The
output current will be displayed in the output current display.
2. Set the desired current value using the rotary Adjust/Enter control.
3. Press the rotary Adjust/Enter control to transition to Fine Adjust
mode.
The set point blinks faster to indicate fine adjust tracking mode.
4. Set the desired current value using the rotary Adjust/Enter control.
5. Once the desired value has been set, press the rotary Adjust/Enter
control to commit the setting.
The SCPI command (s) for these instructions are:
[[:]SOURce[<channel>]]:CURRent[:LEVEl][:IMMediate]
[:AMPLitude]
Important:
The control circuits have been designed to allow you to set the
output voltage up to 105% over the model-rated maximum value. The power
supply will operate within these extended ranges, but full performance to
specification is not guaranteed.
Important:
The maximum and minimum setting values of the output voltage
are limited by the Over Voltage Protection and Under Voltage Protection
settings.
Important:
The control circuits have been designed so that output
current can be set up to 105% over the model-rated maximum values. The power
supply will operate within these extended ranges, but full performance to
specification is not guaranteed.