Calibration and Troubleshooting
Non-isolated Analog Programming Calibration
Prior to this, the main output must be calibrated first.
Important: In calibration commands, when + or – keys are expected, any other key will exit from the calibration mode.
Non-isolated Voltage Monitoring Calibration
Important: For maximum accuracy at a specific APG level. Follow the calibration procedure again but use the intended operating voltage level in steps 3 and 8.
To calibrate the non-isolated voltage monitoring:
1.Disconnect the load and connect the voltmeter to the output.
2.Connect the voltmeter to the
3.Set the Voltage APG level to 4 V. Send the SCPI command:
SYST:COMM:APR:LEV 4
4. Set the input source as Local. Send the SCPI command:
SYST:REM:SOUR LOC
5. Set the main output voltage to 100% by sending the command:
SOUR:VOLT <ModelVolt>
or use the rotary adjust/Enter control.
6.Turn the main output On by pressing the OUTPUT ENABLE Main button on the front panel or send the command:
:OUTP ON
The OUTPUT ENABLE Main button will illuminate.
7. Enter interactive calibration mode by sending the SCPI command:
CAL:MON
8.Press on the terminal keys – or + to reduce or increase monitoring value. Adjust the monitoring value displayed on the voltmeter that was set up in Step 1 to read as close as possible to 4.000 V.
9.Press any key other than the + or – keys to exit from interactive calibration mode.