Analog Programming Mode
M370046-01 4-11
4
Query for Analog Voltage Input Level
To query for analog voltage input level from non-isolated input:
1. Turn the 9-position mode control to the VAP position or press the
rotary Adjust/Enter control if the mode control is already at the VAP
position.
UAPr is displayed on the output voltage display.
2. If the mode is currently nonE then select any other APG mode. If this
step is done, be sure to restore the APG setting to nonE when the
operation is complete. Otherwise, do not rotate the rotary Adjust/Enter
control to keep the current APG mode.
3. Press the rotary Adjust/Enter control to select the APG mode.
The output voltage display should read LE U and the output current
display will display APG level.
4. Press the rotary Adjust/Enter control to return to normal mode.
5. If the APG mode was modified in step 2, execute the procedure to
deactivate the APG voltage mode.
Quick Tip
Remote operation
uses these SCPI
commands. See
page 5–1.
The SCPI commands for these instructions are:
[:]SYSTem[<channel>]:REMote:SOURce[:VOLTage] <?|LOCal
|AVOLtage|IAVoltage|ARESistive|IAResistive>
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel[:VOLTage]
[:ISOlated]
[:]MEASure[<channel>][:SCALar]:APRogram[:VOLTage][:DC]?