Isolated Analog Programming Mode (ISOL)

Resistive-Controlled Voltage ISOL Setup

Activating ISOL Programming Voltage Mode

To activate ISOL programming voltage mode using an external resistor:

1.Turn the 9-position mode control to the VAP position or press the rotary adjust/Enter control if the control knob is already at the VAP position.

UAPr is displayed on the output voltage display.

2.Turn the rotary Adjust/Enter control to select R IS (resistor source from isolated connector).

3.Press the rotary adjust/Enter control to commit the selected mode. LE U is displayed on the output voltage display.

4.Select the input source level using the rotary adjust/Enter control.

5.Once the desired level has been selected, press the rotary Adjust/ Enter control to commit the change.

The display will return to the normal display.

4

Deactivating ISOL Voltage Mode

To deactivate the ISOL voltage mode:

1.Turn the 9-position mode control to the VAP position or press the rotary Adjust/Enter control if the control knob is already at the VAP position.

2.UAPr is displayed on the output voltage display.

3.Turn the rotary Adjust/Enter control to select nonE.

4.Press the rotary adjust/Enter control to commit the change. The display will return to the normal display.