Remote Operation

Selecting the Appropriate Communication Port

Five ports are available for remote digital programming and readback:

RS-232

RS-485

USB

GPIB (optional)

ENET (optional)

To select a communication port:

1.Turn the 9-position mode control to PGM. rE is displayed in the output voltage display.

2.Turn the rotary Adjust/Enter control to select one of the communication ports: 232, 485, gPIb, USb, LAn.

LOCL Loc is also an available option which lets you lock the front panel to prevent the settings from being changed.

Important: The remote interface that was previously configured when the local lock setting was activated will still be available for remote control while the front panel is locked.

Data Rate Setting (Kbps)

The data rate setting is available for RS-232 and RS-485 protocols only. Five optional rates are possible: 1.2, 2.4, 4.8, 9.6 (default), 19.2, 38.4, and

57.6kilo bits per second (Kbps).

To select the desired rate:

1.Turn the 9-position mode control to PGM. rE is displayed in the output voltage display.

2.Turn the rotary Adjust/Enter control to select 232 (RS-232) or 485 (RS-485) communication ports.

3.Press the rotary Adjust/Enter control.

HbPS is displayed on the output voltage display and the currently set data rate is displayed on the output current display.

4.Turn the rotary Adjust/Enter control to select the desired data rate.

5.Press enter to commit the new data rate.

The multichannel address setting will now be configured.