SSI America 9210 user manual General Frequency

Page 58

M4557 - Model 9210 Nitriding Controller

4.1General Frequency

The Nitriding Sample Cell based analyzer is an extremely stable instrument and requires only very occasional calibration. The exact calibration period depends on the type of sample and environment the instrument is placed in. We recommend that any quality assurance procedures written for the instrument are written to allow verification as opposed to calibration. Verification involves checking that the instrument provides the correct analysis of a standard gas within the limits of the instrument and only calibrating when a result outside of limits is produced. The frequency of the verification would need to be in line with the quality regime being operated by the user.

4.2Calibration method

NOTE: Calibration requires a computer (laptop) with an available serial port and HyperTerminal software.

Introduce the calibration gas in the same way as the normal sample.

The sample cell should be isolated from the process gas and the calibration

gas introduced into the Sensor at the correct flow rate. Note that pressure regulators and gauges that may be in the calibration gas lines all have a certain amount of dead space within them and so may require purging for several minutes before the delivered gas matches that of the cylinder contents. Only when the reading is steady has the dead space been purged.

The recommended order of calibration is:

Zero=xxx.xx<CR><LF>.command where xxx.xx = 0 (See appendix 1) The Zero point will use 100% Nitrogen

Span=xxx.xx<CR><LF>. command where xxx.xx = 100 (See appendix 1) The Span point will use 100% Hydrogen

SSi Manual SERIES 9210-M4557-Nitriding

Page 58 of 58

Image 58
Contents SSi’s Nitriding Control System Table of Contents M4557 Model 9210 Nitriding Controller Safety M4557 Model 9210 Nitriding Controller About This Manual Controller DescriptionModel 9210 Terminals Connections Ethernet Connections Flash Card & Flash Card Reader Electrical InstallationOperator Interface Screen Saver Nitriding Control System Mode of Operation Installation Mounting Stand-alone Flow Panel Copper Brass Galvanized Steel AluminumTouch-screen Interface Loops Display Temp DetailMenu Display Menus M4557 Model 9210 Nitriding Controller Program Display Alarm Ack ChartData Logging using Flash Card M4557 Model 9210 Nitriding Controller Program Edit Configuration Configuration MenuParameter Value Date and Time Auxiliary Analog InputShutdown Display Slave Communications StatusManual Event Control PID Loop SetupChange setpoint overshoot protection Event Run Program Setup Zone / Load TC SetupPort Setup Instrument SetupStep Opcode Temperature Atmosphere Option Zone AssignmentsInternal Loop Instrument 1 Instrument Furnace SetupNone BP Back Pressure Temp Internal LoopInternal Instrument 1 Instrument Default Wait LimitsFurnace Name Alarm Setup Relay AssignmentAnalog Input Setup Relay Set PointsAnalog Output Setup Passcode and Alarm IP AddressEvent Control Valve Setup M4557 Model 9210 Nitriding Controller Zero Output Menu Item Security Level Read/Write Raw Data Analog Input Board Setup AUX SP Configuration Programs Overview M4557 Model 9210 Nitriding Controller Series 9210 Opcodes Programmer Description OpcodesM4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller Recipe Step OpCode Temp Applications Information Standard Event AssignmentsTypical Nitriding Instrument Designations OptionsJump Recipe Flow Section Nitriding Gas Supply High and Low Pressure SwitchesVessel Pressure Switches Sample PortsMass Flow Meters Flow Head UnitGas Solenoids Drip Legs Back Pressure ValveDifferential Pressure Transmitter Sample Cell SolenoidM4557 Model 9210 Nitriding Controller System Description Specification Display NA Analog outputs isolated OptionalSensor General Frequency Technical Information Data=xCRLF Error messages System errors Nitriding Sample Cell cleaning TC Type Range in mV E80 J80 K80 N80 NNM Revision History Description Date MCO #