M4557 - Model 9210 Nitriding Controller

DO NOT CHANGE THE FILTER WHILE AMMONIA OR DISASSOCIATED AMMONIA IS FLOWING. THE GASES ARE TOXIC AND VERY HAZARDOUS.

The steps for replacing the filter are as follows;

1.Close all incoming gas valves.

2.Unscrew the clear bowl from the filter housing.

3.Unscrew the filter holder. The holder is the black plastic knob in the center of the filter.

4.Pull off old filter and replace with new.

5.Screw the filter holder back into the top of the filter housing.

6.Screw the clear bowl back on the housing.

7.Open incoming gas lines

Flow to the sample cell should be a minimum of .5 scfh and a maximum of 2.0. This flow is critical to the nitriding process. If flow is low replace the filter. If you are unable to maintain flow contact Super Systems Inc at 1-800-666-4330 immediately.

SSi Manual SERIES 9210-M4557-Nitriding

Page 54 of 54

Image 54
Contents SSi’s Nitriding Control System Table of Contents M4557 Model 9210 Nitriding Controller Safety M4557 Model 9210 Nitriding Controller About This Manual Controller DescriptionModel 9210 Terminals Connections Ethernet Connections Electrical Installation Flash Card & Flash Card ReaderOperator Interface Screen Saver Nitriding Control System Mode of Operation Installation Mounting Stand-alone Flow Panel Copper Brass Galvanized Steel AluminumTouch-screen Interface Loops Display Temp DetailMenu Display Menus M4557 Model 9210 Nitriding Controller Program Display Chart Alarm AckData Logging using Flash Card M4557 Model 9210 Nitriding Controller Configuration Configuration Menu Program EditParameter Value Date and Time Auxiliary Analog InputShutdown Display Slave Communications StatusPID Loop Setup Manual Event ControlChange setpoint overshoot protection Event Run Program Setup Zone / Load TC SetupPort Setup Instrument SetupStep Opcode Temperature Atmosphere Option Zone AssignmentsInternal Loop Instrument 1 Instrument Furnace SetupNone BP Back Pressure Temp Internal LoopDefault Wait Limits Internal Instrument 1 InstrumentFurnace Name Alarm Setup Relay AssignmentAnalog Input Setup Relay Set PointsAnalog Output Setup Passcode and Alarm IP AddressEvent Control Valve Setup M4557 Model 9210 Nitriding Controller Zero Output Menu Item Security Level Read/Write Raw Data Analog Input Board Setup AUX SP Configuration Programs Overview M4557 Model 9210 Nitriding Controller Series 9210 Opcodes Programmer Description OpcodesM4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller Recipe Step OpCode Temp Applications Information Standard Event AssignmentsTypical Nitriding Instrument Designations OptionsJump Recipe Flow Section Nitriding Gas Supply High and Low Pressure Switches Vessel Pressure Switches Sample PortsFlow Head Unit Mass Flow MetersGas Solenoids Drip Legs Back Pressure ValveDifferential Pressure Transmitter Sample Cell SolenoidM4557 Model 9210 Nitriding Controller System Description Specification Display NA Analog outputs isolated OptionalSensor General Frequency Technical Information Data=xCRLF Error messages System errors Nitriding Sample Cell cleaning TC Type Range in mV E80 J80 K80 N80 NNM Revision History Description Date MCO #