SSI America 9210 user manual Nitriding Sample Cell cleaning

Page 63

M4557 - Model 9210 Nitriding Controller

Nitriding Sample Cell cleaning

Safety Notes- Warning!

Hazardous gas or dangerous chemicals formed by the process may be present.

Check with your site safety officer before handling or removing the sensor.

Isolate all external circuits before removing sensor.

Important!

Do not push objects into the sensor, it is easily damaged!

The sensor must be dry before it is reconnected and power applied.

1.Remove sensor from its mounting bracket by slackening the mount fitting (See Figure 1).

2.Disconnect the cable according to connector type:

Terminal: Unscrew the connections ensuring the label specifying color is still present. If removed note the color sequence.

Push fit: Gently pull the latching part to remove the connector.

3.Wash the sensor in de-ionized water by filling the sensor tube and repeatedly flushing. Use gentle agitation for a deeper clean.

4.Drying can be achieved by:

Washing the sensor with iso-propanol (IPA) then immediately blowing dry air or heating (max 60c).

Allowing Nitrogen (dry gas) to flood the cell for a period of 12 hours. Heating (max 60c) can reduce time period.

WARNING: Flammable, do not use on live electrical equipment or other sources of ignition.

5.Refit sensor and allow the reading to settle before following the calibration routine specified in the user manual.

(FIG. 1)

SSi Manual SERIES 9210-M4557-Nitriding

Page 63 of 63

Image 63
Contents SSi’s Nitriding Control System Table of Contents M4557 Model 9210 Nitriding Controller Safety M4557 Model 9210 Nitriding Controller Controller Description About This ManualModel 9210 Terminals Connections Ethernet Connections Electrical Installation Flash Card & Flash Card ReaderOperator Interface Screen Saver Nitriding Control System Mode of Operation Copper Brass Galvanized Steel Aluminum Installation Mounting Stand-alone Flow PanelTouch-screen Interface Temp Detail Loops DisplayMenu Display Menus M4557 Model 9210 Nitriding Controller Program Display Chart Alarm AckData Logging using Flash Card M4557 Model 9210 Nitriding Controller Configuration Configuration Menu Program EditParameter Value Slave Communications Status Auxiliary Analog InputShutdown Display Date and TimePID Loop Setup Manual Event ControlChange setpoint overshoot protection Zone / Load TC Setup Event Run Program SetupInstrument Setup Port SetupZone Assignments Step Opcode Temperature Atmosphere OptionInternal Loop Furnace SetupNone BP Back Pressure Temp Internal Loop Instrument 1 InstrumentDefault Wait Limits Internal Instrument 1 InstrumentFurnace Name Relay Assignment Alarm SetupRelay Set Points Analog Input SetupAnalog Output Setup IP Address Passcode and AlarmEvent Control Valve Setup M4557 Model 9210 Nitriding Controller Zero Output Menu Item Security Level Read/Write Raw Data Analog Input Board Setup AUX SP Configuration Programs Overview M4557 Model 9210 Nitriding Controller Opcodes Series 9210 Opcodes Programmer DescriptionM4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller Options Applications Information Standard Event AssignmentsTypical Nitriding Instrument Designations Recipe Step OpCode TempJump Recipe Sample Ports High and Low Pressure SwitchesVessel Pressure Switches Flow Section Nitriding Gas SupplyFlow Head Unit Mass Flow MetersGas Solenoids Sample Cell Solenoid Back Pressure ValveDifferential Pressure Transmitter Drip LegsM4557 Model 9210 Nitriding Controller System Description Display NA Analog outputs isolated Optional SpecificationSensor General Frequency Technical Information Data=xCRLF Error messages System errors Nitriding Sample Cell cleaning TC Type Range in mV E80 J80 K80 N80 NNM MCO # Revision History Description Date