M4557 - Model 9210 Nitriding Controller

Appendix 1

1.1Communications Protocols.

1.2RS232 Command communications.

Data is transmitted upon request in normal operation and calibration modes.

Transmission is suspended during boot and in edit mode, and in the case of an error. The protocol can be either in a terse format (i.e.D) or in instruments supporting the extended readable format (i.e. Data). Command : DTE to instrument Format : Command=xxx.xx <CR><LF>.

Command

function

reply

 

 

 

Data=x<CR><LF>.

Request for diagnostic

 

 

data. Optional =”x” is

D2 Ref=1234b D1 M1=

 

specifying the required line

2222b

 

of data

 

Reading=x<CR><LF>.

Request for Reading.

R2 CO2=0.01r R1 H2=

 

Optional =”x” is specifying

20.0%

 

the required line of reading

 

 

in sensor with multiple

 

 

data sets.

 

 

 

 

Zero=xxx.xx<CR><LF>.

Request to zero the

Z1 pass or Z1 fail

 

sensor assumed 0.00.

 

 

Optional. =”xxx.xx” is

 

 

specifying the zero point

 

 

data.

 

 

 

 

Span=xxx.xx<CR><LF>.

Request to Span the

S1 pass or S1 fail

 

sensor. data assumed

 

 

100.00% of sensors span.

 

 

=”xxx.xx” is optional

 

 

specifying the Span point

 

 

data. Otherwise.

 

 

 

 

=XXX.xx is optional and must be proceeded by an “=” and is used to set the value of the action. The message is: Reqest Y Quantity = xxxxx unit <CR><LF>.

Request: is the first character of the command D R S Z

Y:is the line number for the data.

The highest line number is always presented first hence line 1 is always the last data line of a communication.

Extension is when “x” is specified the only that line number will be returned.

SSi Manual SERIES 9210-M4557-Nitriding

Page 60 of 60

Image 60
Contents SSi’s Nitriding Control System Table of Contents M4557 Model 9210 Nitriding Controller Safety M4557 Model 9210 Nitriding Controller About This Manual Controller DescriptionModel 9210 Terminals Connections Ethernet Connections Electrical Installation Flash Card & Flash Card ReaderOperator Interface Screen Saver Nitriding Control System Mode of Operation Installation Mounting Stand-alone Flow Panel Copper Brass Galvanized Steel AluminumTouch-screen Interface Loops Display Temp DetailMenu Display Menus M4557 Model 9210 Nitriding Controller Program Display Chart Alarm AckData Logging using Flash Card M4557 Model 9210 Nitriding Controller Configuration Configuration Menu Program EditParameter Value Auxiliary Analog Input Shutdown DisplayDate and Time Slave Communications StatusPID Loop Setup Manual Event ControlChange setpoint overshoot protection Event Run Program Setup Zone / Load TC SetupPort Setup Instrument SetupStep Opcode Temperature Atmosphere Option Zone AssignmentsFurnace Setup None BP Back Pressure TempInternal Loop Instrument 1 Instrument Internal LoopDefault Wait Limits Internal Instrument 1 InstrumentFurnace Name Alarm Setup Relay AssignmentAnalog Input Setup Relay Set PointsAnalog Output Setup Passcode and Alarm IP AddressEvent Control Valve Setup M4557 Model 9210 Nitriding Controller Zero Output Menu Item Security Level Read/Write Raw Data Analog Input Board Setup AUX SP Configuration Programs Overview M4557 Model 9210 Nitriding Controller Series 9210 Opcodes Programmer Description OpcodesM4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller Applications Information Standard Event Assignments Typical Nitriding Instrument DesignationsRecipe Step OpCode Temp OptionsJump Recipe High and Low Pressure Switches Vessel Pressure SwitchesFlow Section Nitriding Gas Supply Sample PortsFlow Head Unit Mass Flow MetersGas Solenoids Back Pressure Valve Differential Pressure TransmitterDrip Legs Sample Cell SolenoidM4557 Model 9210 Nitriding Controller System Description Specification Display NA Analog outputs isolated OptionalSensor General Frequency Technical Information Data=xCRLF Error messages System errors Nitriding Sample Cell cleaning TC Type Range in mV E80 J80 K80 N80 NNM Revision History Description Date MCO #