SSI America 9210 user manual Full Calibration, Menu Item Security Level

Page 36

M4557 - Model 9210 Nitriding Controller

Span Output 2

Enter measured output current (mA)

Calibrate

Edit

<-- Back

 

Skip

 

Next -->

 

DONE

The next screen is the Span Output 2 screen. This screen will allow the user to set the span value for Output 2. Measured at terminals 26(-) and 25(+) for this step. The progress bar at the bottom of the screen will display the calibration progress.

The next screen indicates that the calibration process is complete.

Full Calibration

This screen is used by SSi personnel, and it is a longer version of the User Calibration menu.

Set Menu Security

This menu allows the user to set the security level for all of the menu screens to one of three levels:

1.Operator – Lowest Level, No Pass code Required

2.Supervisor – Middle Level, Level 1 Pass code Required

3.Administrator – Highest Level, Level 2 Pass code Required.

The menu items with security levels are:

Menu Item

Security Level

Program Edit

Supervisor

Auxiliary Instruments

Operator

Auxiliary Analog Input

Operator

Shutdown

Operator

9210 Date and Time

Supervisor

Slave Communications Status

Supervisor

Manual Event Control

Supervisor

PID Loop Setup

Administrator

Event Run Program Setup

Administrator

Zone/Load TC Setup

Administrator

Port Setup

Administrator

Instrument Setup

Administrator

Zone Assignments

Administrator

Furnace Setup

Administrator

Default Wait Limits

Administrator

Furnace Name

Administrator

Alarm Setup

Administrator

Relay Assignments

Administrator

Relay Setpoints

Administrator

Analog Input Setup

Administrator

Analog Output Setup

Administrator

Passcode and Alarm

Administrator

IP Address

Administrator

Event Control

Administrator

Valve Configuration

Administrator

Valve Setup

Administrator

User Calibration

Administrator

SSi Manual SERIES 9210-M4557-Nitriding

Page 36 of 36

Image 36
Contents SSi’s Nitriding Control System Table of Contents M4557 Model 9210 Nitriding Controller Safety M4557 Model 9210 Nitriding Controller About This Manual Controller DescriptionModel 9210 Terminals Connections Ethernet Connections Electrical Installation Flash Card & Flash Card ReaderOperator Interface Screen Saver Nitriding Control System Mode of Operation Installation Mounting Stand-alone Flow Panel Copper Brass Galvanized Steel AluminumTouch-screen Interface Loops Display Temp DetailMenu Display Menus M4557 Model 9210 Nitriding Controller Program Display Chart Alarm AckData Logging using Flash Card M4557 Model 9210 Nitriding Controller Configuration Configuration Menu Program EditParameter Value Auxiliary Analog Input Shutdown DisplayDate and Time Slave Communications StatusPID Loop Setup Manual Event ControlChange setpoint overshoot protection Event Run Program Setup Zone / Load TC SetupPort Setup Instrument SetupStep Opcode Temperature Atmosphere Option Zone AssignmentsFurnace Setup None BP Back Pressure TempInternal Loop Instrument 1 Instrument Internal LoopDefault Wait Limits Internal Instrument 1 InstrumentFurnace Name Alarm Setup Relay AssignmentAnalog Input Setup Relay Set PointsAnalog Output Setup Passcode and Alarm IP AddressEvent Control Valve Setup M4557 Model 9210 Nitriding Controller Zero Output Menu Item Security Level Read/Write Raw Data Analog Input Board Setup AUX SP Configuration Programs Overview M4557 Model 9210 Nitriding Controller Series 9210 Opcodes Programmer Description OpcodesM4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller M4557 Model 9210 Nitriding Controller Applications Information Standard Event Assignments Typical Nitriding Instrument DesignationsRecipe Step OpCode Temp OptionsJump Recipe High and Low Pressure Switches Vessel Pressure SwitchesFlow Section Nitriding Gas Supply Sample PortsFlow Head Unit Mass Flow MetersGas Solenoids Back Pressure Valve Differential Pressure TransmitterDrip Legs Sample Cell SolenoidM4557 Model 9210 Nitriding Controller System Description Specification Display NA Analog outputs isolated OptionalSensor General Frequency Technical Information Data=xCRLF Error messages System errors Nitriding Sample Cell cleaning TC Type Range in mV E80 J80 K80 N80 NNM Revision History Description Date MCO #