| M4557 - Model 9210 Nitriding Controller | |
|
|
|
|
|
| System errors: |
|
| |
|
|
|
| |
| Code | Error | Description |
|
|
|
|
|
|
| NVRAM CRC error | Three areas exist (1,3 & 5) each with a backup (2,4 & 6). Area 1 |
| |
|
|
| contains user parameters i.e. calibration data and the backup is write |
|
|
|
| protected so if this area is restored user calibration etc is LOST. |
|
|
|
| Error 71 will be reported in response to any read request until either |
|
|
|
| a calibration is performed or the instrument is rebooted. Areas 3 & 5 |
|
|
|
| are read only so they can be silently restored. Errors 73 & 75 |
|
|
|
| therefore should never be reported, although they exist internally |
|
|
|
| until they clear themselves. Errors 72,74 & 76 will be reported in |
|
|
|
| response to any read request and cannot be cleared. |
|
|
|
|
|
|
| 77- 78 | TCD curve error | In reading curve data from EEPROM the instrument has attempted |
|
|
|
| to read beyond the ends of the curve. |
|
|
|
|
|
|
| 79 | Wrong block no. | The serial numbers of the Nitriding Sample Cell stored in areas 1 & 5 |
|
|
|
| (in separate devices) do not match |
|
|
|
|
|
|
| 80 | UART missing/error. | For debugging only. Since the only interface available is RS232, this |
|
|
|
| error can never be reported! |
|
|
|
|
|
|
| 81 | Reserved | For external EEPROM not found. |
|
|
|
|
|
|