Agilent Technologies 54622A, 54624A, 54621D, 54622D, 54621A manual General Information, Trigger System

Models: 54621A 54622A 54622D 54621D 54624A

1 111
Download 111 pages 42.09 Kb
Page 18
Image 18
Trigger System

General Information

Trigger System

*Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.

Trigger System

Sources:

54621A/22A: Ch 1, 2, line, ext

 

54621D/22D: Ch 1, 2, line, ext, D15 - D0

 

54624A: Ch 1, 2, 3, 4, line, ext

Modes

Auto, Auto level, Triggered (normal), Single

Holdoff Time

~60 ns to 10 seconds

Selections

Edge, Pattern, Pulse Width, CAN, Duration, I2C, Sequence, SPI, TV, USB

Edge

Trigger on a rising or falling edge of any source.

Pattern

Trigger on a pattern of high, low, and don’t care levels and a rising or falling edge

 

established across any of the sources. The analog channel’s high or low level is defined

 

by that channel’s trigger level.

Pulse Width

Trigger when a positive- or negative-going pulse is less than, greater than, or within a

 

specified range on any of the source channels.

 

Minimum pulse width setting: 5 ns

 

Maximum pulse width setting: 10 s

CAN

Trigger on CAN (Controller Area Network) version 2.0A and 2.0B signals. It can trigger

 

on the Start of Frame bit of a data frame, a remote transfer request frame, or an overload

 

frame.

Duration

Trigger on a multi-channel pattern whose time duration is less than a value, greater

 

than a value, greater than a time value with a timeout value, or inside or outside of a set

 

of time values.

 

Minimum duration setting: 5 ns

 

Maximum duration setting: 10 s

I2C

Trigger on I2C (Inter-IC bus) serial protocol at a start/stop condition, a restart, a missing

 

acknowledge, or user defined frame with address and/or data values. Also trigger on

 

Missing Acknowledge, Restart, EEPROM read, and 10-bit write.

LIN

Trigger on LIN (Local Interconnect Network) sync break at beginning of message frame.

Sequence

Find event A, trigger on event B, with option to reset on event C or time delay.

SPI

Trigger on SPI (Serial Peripheral Interface) a data pattern during a specific framing

 

period. Support positive and negative Chip Select framing ad well as clock Idle framing

 

and user-specified number of bits per frame.

USB

Trigger on USB (Universal Serial Bus) Start of Packet, End of Packet, Reset Complete,

 

Enter Suspend, or Exit Suspend on the differential USB data lines. USB low speed and

 

high speed are supported.

TV

Trigger on any analog channel for NTSC, PAL, PAL-M, or SECAM broadcast standards

 

on either positive or negative composite video signals. Modes supported include Field

 

1, Field 2, or both, all lines, or any line within a field. Also supports triggering on non-

 

interlaced fields. TV trigger sensitivity: 0.5 division of synch signal.

Autoscale

Finds and displays all active analog and digital (for 54621D/54622D) channels, sets edge

 

trigger mode on highest numbered channel, sets vertical sensitivity on analog channels

 

and thresholds on digital channels, time base to display ~1.8 periods. Requires minimum

 

voltage >10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz.