Method 2 — Using the Test Fixture
INP UT S
01660-63801 Test Fixture
STO R A G E
Testing Performance

To verify digital channel threshold accuracy

 

 

b

Method 2 — Using the Test Fixture

 

 

 

Use the Agilent 01660-63801 Performance Verification Test Fixture and the

 

Agilent 01650-61607 cable, BNC Tee, and BNC cable to connect the digital

Figure 3-3

channels D0 - D15 to the oscilloscope

calibrator.￿￿￿￿￿$

See figure 3-3.

 

Oscilloscope Calibrator

1 6 C HAN N EL 5 0 0 MS a/s M ix e d S i g n a l O s c i l lo s c o p e

 

Me a sure tim e

E ntry

 

Sa ve/Rec al l

H O R IZ O N TA L

 

T R IG G ER

 

 

 

 

 

Dela y See figure 3-3.

C H AN N EL

Selec t

Time /Div

Digital Multimeter

HP 34401A

Po sition

 

Trig g e r o ut

Ex t trig g e r in

L i n e

!

!

01650-61607 Cable

BNC-Banana cable

thresh2.cdr

Setting Up Equipment and Test Fixture for the Threshold Test

4Use a BNC-banana cable to connect the multimeter to the other side of the BNC Tee.

5Connect the BNC Tee to the Channel 1 output of the calibrator as shown in figure 3-2 and figure 3-3.

6On the oscilloscope, press the D7 Thru D0 key, then press the Threshold softkey.

3-8