Agilent Technologies 54622D Agilent 54621A/21D Performance Test Record, Agilent 54621A/54621D

Models: 54621A 54622A 54622D 54621D 54624A

1 111
Download 111 pages 42.09 Kb
Page 62
Image 62
Agilent 54621A/21D Performance Test Record

Agilent 54621A/21D Performance Test Record

Agilent 54621A/54621D

Serial No. ______________________________________

Test by _____________________________

Test Interval ____________________________________

Work Order No. ______________________

Recommended Next Testing ________________________

Temperature ____________

 

 

 

 

 

 

 

 

Threshold

Specification

Limits

Ch D7-D0

Ch D15-D8

 

 

Accuracy Test

5 V - 250 mV

4.750 V

________

________

 

 

(100 mV + 3% of

5 V + 250 mV

5.250 V

________

________

 

 

threshold setting)

-5 V - 250 mV

-5.250 V

________

________

 

 

 

-5 V + 250 mV

-4.750 V

________

________

 

 

 

0 V - 100 mV

-100 mV

________

________

 

 

 

0 V + 100 mV

100 mV

________

________

 

 

 

 

 

 

 

 

Voltage Measurement Accuracy

 

 

 

 

 

Range

Power Supply Setting

Test Limits

Channel 1

Channel 2

Channel 3

Channel 4

5 V/Div

35 V

34.04 V to 35.96 V

________

________

________

________

2 V/Div

14 V

13.616 V to 14.384 V

________

________

________

________

1 V/Div

7 V

6.808 V to 7.192 V

________

________

________

________

500 mV/Div

3.5 V

3.404 V to 3.596 V

________

________

________

________

200 mV/Div

1.4 V

1.3616 V to 1.4384 V

________

________

________

________

100 mV/Div

700 mV

680.8 mV to 719.2 mV

________

________

________

________

50 mV/Div

350 mV

340.4 mV to 359.6 mV

________

________

________

________

20 mV/Div

140 mV

136.16 mV to 143.84 mV

________

________

________

________

10 mV/Div

70 mV

68.08 mV to 71.92 mV

________

________

________

________

5 mV/Div

35 mV

34.04 mV to 35.96 mV

________

________

________

________

2 mV/Div

14 mV

13.616 mV to 14.384 mV

________

________

________

________

1 mV/Div

7 mV

6.616 mV to 7.384 mV

________

________

________

________

 

 

 

 

 

 

 

Bandwidth

 

Test Limits

Channel 1

Channel 2

 

 

 

 

3 dB at 60 MHz

________

________

 

 

 

 

 

 

 

 

Horizontal t and 1/t Accuracy

 

 

 

 

 

 

Generator Setting

Test Limits

Results

 

 

 

Frequency

10 kHz

9.98 kHz to 10.02 kHz

________

 

 

 

Period

100 s

99.79 s to 100.2 s

________

 

 

 

Frequency

1 MHz

997.9 kHz to 1.002 MHz

________

 

 

 

Period

1 s

997.9 ns to 1.002 s

________

 

 

 

Frequency

50 MHz

49.77 MHz to 50.23 MHz

________

 

 

 

Period

20 ns

19.91 ns to 20.09 ns

________

 

 

 

 

 

 

 

 

 

 

Trigger Sensitivity

 

Test Limits

Channel 1

Channel 2

 

 

 

Internal trigger

25 MHz at 0.35 division

________

________

 

 

 

 

60 MHz at 0.35 division

________

________

 

 

 

 

 

Ext

 

 

 

 

External trigger

25 MHz at <75 mVp-p

________

 

 

 

 

 

60 MHZ at <150 mVp-p

________

 

 

 

3-21

Page 62
Image 62
Agilent Technologies 54622D, 54624A, 54622A manual Agilent 54621A/21D Performance Test Record, Agilent 54621A/54621D