C A U T I O N
Figure
Troubleshooting
To check the Low Voltage Power Supply
To check the Low Voltage Power Supply
1Disconnect the power cord from the oscilloscope. Then remove the oscilloscope cover and set the oscilloscope on its side.
USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS !
If you operate the oscilloscope with the cabinet removed, you must use an external fan to blow air across the system board. This external air flow is necessary to cool the heat sinks on the system board. Otherwise, damage to the components can occur.
2 | ||||
Connect the negative lead of the multimeter to a ground point on the | ||||
| 9(9( | |||
| oscilloscope. Connect the power cord and turn on the oscilloscope. | |||
3 | Measure the power supply voltages at E1 through E6 on the system | |||
| board. SeeFigure | |||
| 99 |
|
|
|