Troubleshooting

To construct your own dummy load

To construct your own dummy load

1Obtain a connector that is compatible with the connector J3 on the Low Voltage Power Supply.

2Connect load resistors to the connector as follows:

Connector

Load (Amps)

Resistor

Watts

Connect to Pin

Ground to Pin

+5.1 V

4.4 A

1.2

22 W

10, 11, 12

13, 14

-5.2 V

1.6 A

3.25

8 W

4, 5

6

+15.75 V

1.2 A

13.1

20 W

1

2

+3.3 V

4 A

0.825

13 W

15, 16, 17, 18

19, 20

Figure 5-1

+5.1 / -5.2 V Balance Adjustment

 

+3.3V Adjustment

 

E2 and E3

 

8.5 V DC

Fan Connector

Printer Power

Location of the Low Voltage Power Supply Connectors