Testing Performance
To verify digital channel threshold accuracy
To verify digital channel threshold accuracy
This test verifies the digital channel threshold accuracy specification of the
Agilent 54621D/22D
Test Threshold Accuracy only on the 54621D/22D
You need to perform these instructions only if you will be testing the Agilent 54621D/22D
Threshold accuracy test limits= ±(100 mV + 3% of threshold setting)
When to Test You should perform this test every 24 months or after 4000 hours of operation, whichever comes first.
What to Test Use these instructions to test the threshold settings of digital channels
Verifying Test Results After each threshold test, record the voltage reading in the Performance Test Record at the end of this chapter. To verify whether a test passes, verify that the voltage reading is within the limits in the Performance Test Record.
Table | Equipment Required to Test Threshold Accuracy |
| |
|
|
|
|
| Equipment | Critical Specifications | Recommended Model/Part |
| Digital Multimeter | 0.1 mV resolution, 0.005% | Agilent 34401A |
|
| accuracy |
|
| Oscilloscope Calibrator | DC offset voltage 6.3 V | Fluke 5820A |
|
| Agilent | |
| BNC Tee |
| Agilent |
| BNC Cable |
| Fluke 50Ω cable, P/N 686318 |
| BNC Test Connector, |
| |
|
|
| connector and an |
|
|
| of Berg strip.” on page |
| Test Fixture | PV test fixture | Agilent |
| Probe Cable |
| Agilent |