8Is the fan running?

Troubleshooting

To check out the oscilloscope

8Is the fan running?

The low voltage power supply has a thermal cut-out circuit. If the fan is defective, the low voltage power supply will shut down when it gets too hot for safe operation.

If the fan is running, perform the internal self-tests. Go to “To run the internal self-tests” on page 5-13.

If the fan is not running, it may be defective. Follow these steps: a Disconnect the fan cable from the power supply.

b Measure the fan voltage at the connector on the power supply. See Figure 5-2 for the location of the fan connector.

c If the fan voltage is approximately +8 Vdc, replace the fan. If the fan voltage is not approximately +8 Vdc, replace the power supply.

Figure 5-2

+3.3V Adjustment

+5.1 / -5.2 V

Fan Connector

Printer Power

Balance Adjustment

Location of the Fan Connector