To adjust the power supply

Calibrating and Adjusting

To adjust the power supply

To adjust the power supply

The oscilloscope power supply has both a +3.3 V adjustment, and a

+5.1 V / -5.2 V balance adjustment. Other oscilloscope voltages are based on the +3.3 V adjustment.

In this procedure, you will use a digital multimeter to measure the +3.3 V and +5.1 V / -5.2 V test points, and adjust the power supplies to be within tolerance, if necessary.

Table 4-1

Equipment Required to Adjust the Power Supply

 

 

 

 

 

 

Equipment

Critical Specifications

Recommended Model/Part

 

Digital multimeter

0.1 mV resolution, accuracy ±0.05%

Agilent 34401A

1Prepare the oscilloscope for the voltage adjustment.

a Turn off the oscilloscope and disconnect the power cable.

b Remove the oscilloscope cover. For a list of parts, see the “Replaceable Parts” chapter, then return here.

c Place the oscilloscope on its side.

d Connect the negative lead of the digital multimeter to a ground point on the oscilloscope chassis.

e Reconnect the power cable. f Turn on the oscilloscope.