Agilent Technologies 54622A, 54624A, 54621D Wait a few seconds for the measurement to settle, 3-12

Models: 54621A 54622A 54622D 54621D 54624A

1 111
Download 111 pages 42.09 Kb
Page 53
Image 53
Wait a few seconds for the measurement to settle.

Testing Performance

To verify voltage measurement accuracy

3Use the BNC tee and cables to connect the oscilloscope calibrator /power supply to both the oscilloscope and the multimeter.

4Adjust the output so that the multimeter reading displays the first Volts/div supply setting value in Table 3-5.

Wait a few seconds for the measurement to settle.

5Press the Y2 softkey, then position the Y2 cursor to the center of the voltage trace using the Entry knob.

The Y value on the lower line of the display should be within the test limits of Table 3-5. If a result is not within the test limits, see the “Troubleshooting” chapter. Then return here.

6Continue to check the voltage measurement accuracy with the remaining Volts/div setting values in Table 3-5.

7When you are finished checking all of the power supply setting values, disconnect the power supply from the oscilloscope.

8Repeat this procedure for Channels 2, 3, and 4, if applicable on your oscilloscope model.

Use a Blocking Capacitor to Reduce Noise

On the more sensitive ranges, such as 1 mV/div, 2 mV/div, and 5 mV/div, noise may be a factor. To eliminate the noise, use a BNC Tee, blocking capacitor, and BNC shorting cap to shunt the noise to ground. See figure 3-4.

Figure 3-4

To Power Supply or

Blocking

Capacitor

Calibrator

 

BNC shorting Using a Blocking Capacitor to Reduce Noise3-12Manual background cap

To oscilloscope input

Using a Blocking Capacitor to Reduce Noise

3-12

Page 53
Image 53
Agilent Technologies 54622A, 54624A, 54621D, 54622D, 54621A manual Wait a few seconds for the measurement to settle, 3-12