Testing Performance

To verify bandwidth

To verify bandwidth

This test verifies bandwidth. In this test you will use an oscilloscope calibrator with a level sinewave output.

You will use the peak-to-peak voltage both at 1 MHz and at bandwidth frequency to verify the bandwidth response of the oscilloscope.

54622A, 54622D, and 54624A Test limits at 1 mV/div to 5 V/div:

• All channels (±3 dB)

• dc to 100 MHz

• ac coupled 10 Hz to 100 MHz

54621A and 54621D

Test limits at 1 mV/div to 5 V/div:

• All channels (±3 dB)

• dc to 60 MHz

• ac coupled 10 Hz to 60 MHz

Table 3-6

Equipment Required to Verify Bandwidth

 

 

 

 

 

 

Equipment

Critical Specifications

Recommended Model/Part

 

Oscilloscope Calibrator

 

Fluke 5820A

 

Cable *

Type N (m), 24-inch

Agilent 11500B

 

Feedthrough

50, BNC (m) and (f)

Agilent 11048C

*The oscilloscope calibrator is supplied with 2 or more coaxial cables N (m), BNC (m), 1 meter long, Fluke P/N 686318.

3-13

Page 54
Image 54
Agilent Technologies 54621A, 54624A, 54621D, 54622D, 54622A manual To verify bandwidth, Equipment Required to Verify Bandwidth