Low Voltage Power Supply Adjustments (on the top of the oscilloscope)

Calibrating and Adjusting

To adjust the power supply

4If the +5.1 V measurement is out of tolerance, adjust the

+5.1 V / -5.2 V balance adjustment on the power supply. See Figure 4-2.

5If the -5.2 V measurement is out of tolerance, adjust the

+5.1 V / -5.2 V balance adjustment on the power supply. See Figure 4-2.

The +15.75 V supply is not adjustable, and is dependent upon the +5.1 V supply.

Figure 4-2

+5.1 / -5.2 V Balance Adjustment

 

+3.3V Adjustment

 

E2 and E3

 

8.5 V DC

Fan Connector

Printer Power

Low Voltage Power Supply Adjustments (on the top of the oscilloscope)

6If adjusting the power supply does not bring all the voltages within tolerance, go to the “Troubleshooting” chapter.