Testing Performance
To verify voltage measurement accuracy
bSet the Volts/Div setting to the value in the first line in Table
Table | Settings Used to Verify Voltage Measurement Accuracy | ||||
|
|
|
|
|
|
| Volts/Div Setting | Power Supply Setting | Test Limits |
|
|
| 5 V/Div | 35 V | 34.04 V | to | 35.96 V |
| 2 V/Div | 14 V | 13.616 V | to | 14.384 V |
| 1 V/Div | 7 V | 6.808 V | to | 7.192 V |
| 0.5 V/Div | 3.5 V | 3.404 V | to | 3.596 V |
| 0.2 V/Div | 1.4 V | 1.3616 V | to | 1.4384 V |
| 0.1 V/Div | 700 mV | 680.8 mV | to | 719.2 mV |
| 50 mV/Div | 350 mV | 340.4 mV | to | 359.6 mV |
| 20 mV/Div | 140 mV | 136.16 mV | to | 143.84 mV |
| 10 mV/Div | 70 mV | 68.08 mV | to | 71.92 mV |
| 5 mV/Div | 35 mV | 34.04 mV | to | 35.96 mV |
| 2 mV/Div | 14 mV | 13.616 mV | to | 14.384 mV |
| 1 mV/Div* | 7 mV | 6.616 mV | to | 7.384 mV |
*Full scale is defined as 16 mV on the 1 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
cPress the Acquire key. Then press the Averaging softkey and set #Avgs to 64. Wait a few seconds for the measurement to settle.
2Press the Cursors key, set the Mode softkey to Normal, then press the X Y softkey and select Y. Press the Y1 softkey, then use the Entry knob
(labeled on the front panel) to set the Y1 cursor on the baseline of the signal.