bSet the Volts/Div setting to the value in the first line in Table 3-5.

Testing Performance

To verify voltage measurement accuracy

bSet the Volts/Div setting to the value in the first line in Table 3-5.

Table 3-5

Settings Used to Verify Voltage Measurement Accuracy

 

 

 

 

 

 

 

Volts/Div Setting

Power Supply Setting

Test Limits

 

 

 

5 V/Div

35 V

34.04 V

to

35.96 V

 

2 V/Div

14 V

13.616 V

to

14.384 V

 

1 V/Div

7 V

6.808 V

to

7.192 V

 

0.5 V/Div

3.5 V

3.404 V

to

3.596 V

 

0.2 V/Div

1.4 V

1.3616 V

to

1.4384 V

 

0.1 V/Div

700 mV

680.8 mV

to

719.2 mV

 

50 mV/Div

350 mV

340.4 mV

to

359.6 mV

 

20 mV/Div

140 mV

136.16 mV

to

143.84 mV

 

10 mV/Div

70 mV

68.08 mV

to

71.92 mV

 

5 mV/Div

35 mV

34.04 mV

to

35.96 mV

 

2 mV/Div

14 mV

13.616 mV

to

14.384 mV

 

1 mV/Div*

7 mV

6.616 mV

to

7.384 mV

*Full scale is defined as 16 mV on the 1 mV/div range.

Full scale on all other ranges is defined as 8 divisions times the V/div setting.

cPress the Acquire key. Then press the Averaging softkey and set #Avgs to 64. Wait a few seconds for the measurement to settle.

2Press the Cursors key, set the Mode softkey to Normal, then press the X Y softkey and select Y. Press the Y1 softkey, then use the Entry knob

(labeled Settings Used to Verify Voltage Measurement Accuracy on the front panel) to set the Y1 cursor on the baseline of the signal.

3-11

Page 52
Image 52
Agilent Technologies 54622D, 54624A Testing Performance, To verify voltage measurement accuracy, 3-11, Volts/Div Setting