To verify voltage measurement accuracy

Testing Performance

To verify voltage measurement accuracy

To verify voltage measurement accuracy

This test verifies the voltage measurement accuracy. In this test, you will measure the output of a power supply using dual cursors on the oscilloscope, and compare the results with the multimeter reading.

Test limits: ±2.0% of full scale ±1 LSB*

Full scale is defined as 16 mV on the 1 mV/div range.

Full scale on all other ranges is defined as 8 divisions times the V/div setting.

*1 LSB = 0.4% of full scale

Table 3-4

Equipment Required to Verify Voltage Measurement Accuracy

 

 

 

 

 

Equipment

Critical Specifications

Recommended Model/Part

 

Power supply

14 mV to 35 Vdc,

Fluke 5820A or Agilent 3245A

 

 

0.1 mV resolution

 

 

Digital multimeter

Better than 0.01% accuracy

Agilent 34401A

 

Cable

BNC, Qty 2

Agilent 10503A

 

Shorting cap

BNC

Agilent 1250-0774

 

Adapter

BNC (f) to banana (m)

Agilent 1251-2277

 

Adapter

BNC tee (m) (f) (f)

Agilent 1250-0781

 

Blocking capacitor

 

Agilent 10240B

Do this procedure first for Channel 1. Then repeat the procedure for Channel 2.

1Set up the oscilloscope.

a Adjust the channel 1 position knob to place the baseline at approximately

0.5division from the bottom of the display.

3-10