If you cannot see a channel

Troubleshooting

If you cannot see a channel

If you cannot see a channel

Check that the oscilloscope probe cable is securely connected to the input connector.

Check that the oscilloscope probe lead wires are securely inserted into the connector assembly and that the probe clips make good contact with the probe lead wires.

Check that the probe clips are securely connected to points in the circuit under test.

Check that the circuit under test is powered on.

You may have pressed the Autoscale key before an input signal was available.

Performing the checks listed here ensures that the signals from the circuit under test will be seen by the oscilloscope. Perform the remaining checks in this topic to make sure the oscilloscope channels are on, and to obtain an automatic setup.

Check that the desired oscilloscope channels are turned on.

a Press the analog channel key until it is illuminated.

b Press the digital channel group key (D15 Thru D8 or D7 Thru D0 keys) until it is illuminated. Use the Entry knob or Channel Select knob to scroll through the digital channels to make sure that the desired channels are turned on.

Press the Autoscale key to automatically set up all channels.