BEWARE OF HAZARDOUS VOLTAGES !

Troubleshooting

To check the Low Voltage Power Supply

Table 5-4

Low Voltage Power Supply Tolerances

 

 

 

 

Supply Voltage

Tolerance

 

+5.1 V

±153 mV (+4.947 V to +5.253 V)

 

-5.2 V

±156 mV (-5.04 V to -5.36 V)

 

+15.75 V

+1.260 V, -787 mV (+14.963 V to +17.010 V)

 

+3.3 V

±100 mV (+3.20 V to +3.40 V)

4If the +5.1 V measurement is out of tolerance, adjust the +5.1 V adjustment on the power supply.

5If the -5.2 V measurement is out of tolerance, adjust the -5.2 V adjustment on the power supply.

The +15.75 V supply is not adjustable, and is dependent upon the +5.1 V supply.

If the Fuse is Blown

If the power supply fuse is blown, the power supply is defective, and you must replace it. See the “Replaceable Parts” chapter for information about removing the power supply.

C A U T I O N

BEWARE OF HAZARDOUS VOLTAGES !

 

Be careful when performing component-level repair. Voltages greater than

 

300 Vdc exist, and can cause injury.

 

 