Replacing Assemblies

To remove the cabinet

To remove the cabinet

1Turn off the oscilloscope and disconnect the power cable.

2 Rotate the handle up as shown below.

3 If a module is installed, remove it from the oscilloscope.

4Using the T20 TORX driver, remove the two screws from the rear of the cabinet.

5Using your thumbs, gently push on the rear-panel connectors to slide the oscilloscope out of the cabinet.

Figure 6-1

Removing the cabinet

6-4

Page 88
Image 88
Agilent Technologies 54622A, 54624A, 54621D, 54622D, 54621A manual To remove the cabinet, Removing the cabinet